IEC 60884-1 Clause 30.1 Figure 41 | IEC 60811-3-1 Clause 8 | High Temperature Pressure Test Device

IEC 60884-1 Clause 30.1 Figure 41 | IEC 60811-3-1 Clause 8 | High Temperature Pressure Test Device

Product No: GNGPL-3610-2PA

Leave a Message






    captcha

    Description

    The GNGPL-3610-2PA High Temperature Pressure Test Device is a precision instrument designed for evaluating the thermal stability of insulation and sheathing materials in accordance with international standards. This apparatus applies a calibrated force to test specimens using standardized blades under controlled elevated temperature conditions, followed by precise measurement of indentation depth to quantify material deformation. The resulting data provides objective validation of structural integrity under thermal stress, supporting quality assurance protocols for electrical components and cable assemblies.

    Compliance Standards:

    • IEC 60884-1:2022 – Plugs and socket-outlets for household and similar purposes – Part 1: General requirements
    • IEC 60811-3-1:2020 – Electric and optical cables – Test methods for non-metallic materials – Part 3-1: Mechanical tests – Compression test at elevated temperature
    • GB/T 2099.1-2021 – Household and similar electrical accessories – Plugs and socket-outlets – Part 1: General requirements
    • GB/T 2951.31-2008 – Common test methods for insulating and sheathing materials of electric and optical cables – Part 31: Methods specific to PVC compounds – Pressure test at high temperature

    Technical Specifications:

    Model GNGPL-3610-2PA GNGPL-3610-3PA
    Working Stations 2 stations 3 stations
    Blade Configuration 1 rectangular blade + 1 Φ6mm round blade 3 rectangular blades
    Blade Thickness 0.7mm ±0.01mm
    Applied Load 2.5N standard weights 1-1kg adjustable weights (3 sets)
    Test Temperature 200°C standard (configurable per cable specification when used with LISUN GW-225 High Temperature Chamber)
    Test Duration 2 hours standard 4 hours (D≤15mm) / 6 hours (D>15mm)
    Cooling Transition 10 seconds (rapid cooling via water spray method)
    Acceptance Criteria Indentation depth ≤50% of specimen thickness Median indentation depth ≤50% of average specimen thickness

    Standard Test Configurations:

    • IEC 60884-1 Clause 30.1 Figure 41 – Test apparatus for pressure test at high temperature
    • IEC 60811-3-1 Clause 8 Figures 1, 2 and 3 – Compression test apparatus for elevated temperature testing

    Tags: GNGPL-3610-2PA, High Temperature Pressure Test, Insulation Testing, Cable Testing

    Frequently Asked Questions

    Where is LISUN based and what is your global reach?

    We are based in Shanghai, China, founded in 2012. Our products are sold to Western Europe (25%), North America (20%), Eastern Europe (15%), South America (10%), Domestic Market (5%), and other regions. We have 51–100 employees.

    How do you guarantee product quality?

    We always provide a pre-production sample before mass production and conduct a final inspection before shipment.

    What products can I buy from LISUN?

    Goniophotometer, Integrating Sphere, Temperature Chamber, EMI Receiver, Surge Generator, and more.

    Why should I choose LISUN over other suppliers?

    LISUN is a leader in lighting and electrical test instruments, founded in 2003. All products are ISO9001:2015 certified and CE certified. LISUN is a CIE member, and our products have been sold to over 150 countries.

    What services and payment terms do you offer?

    Delivery terms: FOB, CFR, CIF, EXW, DDU. Payment currencies: USD, EUR, CNY. Payment types: T/T, L/C, Credit Card, PayPal, Cash, Escrow. Languages: English, Chinese, Spanish, Arabic, Russian.